We ship via UPS and fastest delivery is with Next Day Air. UPS Next Day Air Early typically gets their product to them as early as 8am, Next Day Air gets their product typically there by 10:30am and Next Day Air Saver typically by 3pm. We ship same day for in stock items and the lead time is up to one week for out of stock items. Delivery is dependent on UPS and your shipping method.
ADD TO CART
ADD TO CART
BUY BY 1-CLICK

Features

  • Able to record CO, CO2, HCHO, TVOC, Temperature, Humidity and Dew Point Measurements
  • Non-dispersive Infrared (NDIR) CO2 Sensor
  • HCHO : Electrochemical Sensor
  • VCO : Metal-Oxide Sensor
  • Up to 450,000 records with date and time stamps
  • Supports the DCON, Modubs RTU/TCP, and MQTT Protocols
  • Includes RS-485/Ethernet Communication Interfaces
  • Relay Output for Alarm or IAQ Device Control

Introduction

The CL-200 series of data logger devices can be used to record HCHO, TVOC, CO, CO2, temperature, humidity and dew point information, including the date and time stamps, and are able to store up to 450,000 downloadable records.

Real-time data can be accessed from the CL-200 data logger from anywhere and at any time using the free Windows software, the iSO App, or the Android App, as long as they are connected to the same local network as the data logger.

Support is provided for common industrial protocols such as DCON, Modbus RTU, and Modbus TCP, as well as the emerging machineto-machine (M2M)/)IoT (Internet of Things) connectivity protocol – MQTT. The CL-200 data logger can be connected via widely used communication interfaces including RS-485, Ethernet and PoE, meaning that the device can be easily integrated into existing HMI or SCADA systems, ensuring trouble-free maintenance in distributed control systems.

  • Supports the MQTT Protocol for IoT Applications
  • Multi-platform Remote Access Software

    Real-time data from the CL-200 data logger can be accessed from anywhere and at any time using the DL300 Utility, the iOS or Android App, or via a regular web browser, as long as they are connected to the same local network as the data logger.

Simple and Powerful Dl-300 Utility

The DL-300 Utility is a powerful tool that is designed for configuring the modules, monitoring real-time data, grouping DL-300 and CL-200 modules to view and manage the status of distribution groups, downloading log data, which can be exported to a CSV file that can then be imported into any industry-standard software or spreadsheet for analysis.

Applications

  • Transportation of Food or Pharmaceuticals
  • Food and Beverage Industry (HACCP)
  • Blood Stations and Pharmacies
  • Building and Energy Management
  • Warehouse Management
  • Museums, Archives, and Galleries

HCHO (Formaldehyde) Detector Automation Solution
The HCHO detector provides an instantaneous warning of the presence of toxic formaldehyde, which is commonly found in plywood factories, furniture factories, and newly remodelled homes.

Indoor or Underground Parking Lot Automatic Monitoring Solution
An air quality automation system for indoor or underground parking lots can use the Cl-200 series modules to monitor for health and safety indicators such as CO, CO2, temperature, humidity, dew point, and more..

Specifications

System
Model
CL-201-E
CL-202-E
CL-203-E
CL-204-E
CL-207-E
CL-208-E
CO/CO2/HCHO/RH/T Alarm
Yes
Real Time Clock
Yes
Data Logger
Yes, up to 450,000 records
Relay Output
Form A×1, SPST
30 VDC @ 16 A or 250 VAC @ 16 A
Interface
RS-485/Ethernet
Electrical
Powered from Terminal Block
+12 to +48 VDC
Power Consumption
1.6 W (Max.)
1.6 W (Max.)
1.7 W (Max.)
2.0 W (Max.)
2.0 W (Max.)
2.0 W (Max.)
Mechanical
Installation
Ceiling Mounting/Wall Mounting
Protection Class
IP20
Dimensions (D x H)
Ø 150 mm x 53 mm
Environment
Operating Temperature
0 to +50°C
Storage Temperature
-1 ppm
Humidity
10 to 90% RH, Non-condensing
 

I/O Specifications

Model CL-201-E CL-202-E CL-203-E CL-204-E CL-207-E CL-208-E
HCHO Measurement
Range - 0 ppb to 2000 ppb (Electrochemical) 0 ppb to 2000 ppb (Electrochemical) -
Resolution - 1 ppb 1 ppb -
Accuracy - 0 ~ 300ppb: ±30ppb
> 300ppb: ±10%
0 ~ 300ppb : ±30ppb
> 300ppb : ±10%
-
Response Time - ≤60 seconds ≤60 seconds -
  - 180 seconds 180 seconds -
TVOC Measurement
Range - - - 0 ppb to 60000 ppb (MEMS Metal Oxide) - 0 ppb to 60000
ppb (MEMS
Metal Oxide)
Resolution - - - 1 ppb - 1 ppb
Accuracy - - - ±15% - ±15%
Response Time - - - 60 seconds - 60 seconds
Warm-up Time - - - 180 seconds - 180 seconds
CO Measurement
Range 0 to 1000 ppm (Electrochemical) - 0 to 1000 ppm
(Electrochemical)
-
Resolution 1 ppm - 1 ppm -
Accuracy ±5% of measured value - ±5% of measured value -
Response Time 30 seconds - 30 seconds -
Warm-up Time 300 seconds - 300 seconds -
CO2 Measurement
Range - 0 to 9999 ppm (NDIR) -
Resolution - 1 ppm -
Accuracy - ±40 ppm ±3% of measured value -
Response Time - 120 seconds -
Warm-up Time - 5 minutes -
Temperature Measurement
Range -10 to +50°C
Resolution 0.1°C
Accuracy ±0.6°C
Relative Humidity Measurement
Range 0 to 100% RH, Non-condensing
Resolution 0.1% RH, Non-condensing
Accuracy ±5% RH, Non-condensing
Dew Point
Range Calculated using temperature and relative humidity
Resolution 0.1°C